MENU CONFIGURATION DISPLAYS

Not all menu items display on standard meters.
Table 6-1Configuration Menu

MENU
/TARE
DESCRIPTION
INPT
Input Type
Show input choices:
100m
100mVdc
▒50m
50mVdc
10 V 10Vdc
▒5 V 5Vdc
0-20 0-20mAdc
DEC.P
Decimal Point
Show current decimal point position.
FFFF.
FFF.F
FF.FF
F.FFF

RD.S.O (Reading Scale & Offset)

Enter new value and show "RD 1"

Enter new value and show "IN 2"
Show IN 1 and prior value entered.

Scroll to the next digit to the right

Show prior value entered.

Scroll to the next digit to the right.

Show RD 2 and prior value entered.

Scroll to the next digit to the right.



Change the value of the flashing digit.




Change the value of the flashing digit.




Change the value of the flashing digit.
RD.CF
(Reading Configuration)
R.1=


R.2=


















R.3=


R.4=
R (Ratiometric Reading)
N (Non-ratiometric reading)

0 (1╡V resolution for unipolar & 5╡V resolution for bipolar)

1 (2╡V resolution for unipolar & 10╡V resolution for bipolar)

2 (3╡V resolution for unipolar & 15╡V resolution for bipolar)

3 (5╡V resolution for unipolar & 25╡V resolution for bipolar)

4 (10╡V resolution for unipolar & 25╡V resolution for bipolar)

Note: 3 ╡V resolution means if your input is 0-30mV, at 30mV the display shows 9999.

F (Filtered value)
U (Unfiltered value)


G (Gross/Net Display)
P (Peak Display)
S1.CF
Setpoint 1 Configuration
S.1=


S.2=


S.3=
A (Active above)
B (Active below)

U (Unlatched)
L (Latched)

N (Net Reading)
G (Gross Reading)
S2.CF
Setpoint 2 Configuration
S.1=


S.2=


S.3=
A (Active above)
B (Active below)

U (Unlatched)
L (Latched)

N (Net Reading)
G (Gross Reading)
S1.DB
Setpoint 1 Deadband
Press to scroll to the next digit to the right.
Press to change the value of the flashing digit
S2.DB
Setpoint 1 Deadband
Press to scroll to the next digit to the right.
Press to change the value of the flashing digit
OT.CF
Output Configuration
0.1


0.2


0.3
E (Analog output is enabled)
D (Analog output is disabled)

C (Analog output is current)
V (Analog output is voltage)

N (Net Reading)
G (Gross Reading)
Notes: 1) if 0.2 = V, then you may select your analog output to be 0-10 V or 0-5 V by accessing submenu 0.5
OT.S.O
OUTPUT SCALE AND OFFSET

Enter new value and show "OUT1"


Enter new value and show "RD 2"


Enter new value and show "OUT2"
Show RD1 and prior value entered

Scroll to the next digit to the right

Show prior value entered.

Scroll to the next digit to the right

Show prior value entered.

Scroll to the next digit to the right



Change the value of the flashing digit




Change the value of the flashing digit



Change the value of the flashing digit
LK.CF
LOCK OUT CONFIGURATION

RS=




SP=


L3=
E (Enable RESET button in the run mode)
D (Disable RESET button in the run mode)

E (Enable setpoint changes)
D (Disable setpoint changes)

0 (SETPTS button display setpoint values)
1 (SETPTS button display software ôv.-03ö version)

RUN MODE
DISPLAYS
Display
/TARE
/NT/GRS
RESET
DESCRIPTION
NET/GROS
Press to activate
Displays "NET or GROSS" reading.Once reading shows, respective value shows.
 

Peak Reading


Toggles between Net and Gross values.

T.RST
 
Will reset your tare when viewing this function
Tare Reset
SP.RS
      Reset Latched Alarms

Pressing the RESET button resets your latched alarm

WARNING! Tare resets if Tare is activated.